Model 9903

SPECIFICATION

Dimensions
Internal: 38.75″W x 26″D x 21″H (98cm x 66xm x 53xm)External: 52″W x 31″D x 22″H (132cm x 79cm x 56cm)
Construction
Chamber: 3/8″ thick clear & white acrylicEquipment door front: 32″ x 14″ (81 x 35 cm) openingLeft side door: 1/2″ (12mm) clear acrylic with 1/2-turn latch
Hardware Ports
8″ ports with gloves (optional)1×2.25″ (57mm) Cable pass through
Controllers
Two (2) CAL 9500, multiple ramp/soak cycles, Resolution: 0.1%/0.1°
Sensor Accuracy
RH: 2.0% RH at 68°F (20°C) and 0-90% RH, ± 3.0% RH at 68°F (20°C) and 90-100% RHTemp: ± 0.4° F (± 0.2° C) at 20°C
Power
Voltage: 115/230 VAC 50/60HzCurrent: 8Amp/4Amps

Model 9910

SPECIFICATION

Dimensions
17”W x 5”H x 12”D
Construction
Metal enclosure, docks atop 9910 PinPulse.
Weight
23 lbs.
Interface
Waveform connection: Cables supplied for all supported ESD models
• HBM Cable Pair, Red, Black, 6” (152mm)
• MM Cable Pair, Yellow, Green, 6” (152mm)
• HMM Cable Pair, Blue, Orange, 6”(152mm)
Device connection: DUT stage providing connection to 128 circuits.
PinScan interface is dual 64-pin ZIF sockets, .100” pin spacing.
ON/OFF
Rocker switch on front panel Indicators: Red & Green LEDs on front panel indicate status.
Communications
USB connector on rear panel.
Power
90-260 VAC, 50-60 Hz, via IEC connector on rear panel

Model 5533

SPECIFICATION

Dimensions
Internal: 38.75” W x 26” D x 21” H (98cm x 66xm x 53xm)
External: 52” W x 31” D x 22” H (132cm x 79cm x 56cm)
Construction
Chamber: 3/8” thick clear & white acrylic
Equipment door front: 32” x 14” (81 x 35 cm) opening
Left side door: ½” (12mm) clear acrylic with ½-turn latch
Hardware Ports
8” ports with gloves (optional) 1×2.25” (57mm) Cable pass through
Controllers
Two (2) CAL 9500, multiple ramp/soak cycles, Resolution: 0.1%/0.1°
Sensor Accuracy
RH: 2.0% RH at 68°F (20°C) and 0-90% RH , ± 3.0% RH at 68°F (20°C) and 90-100% RH Temp: ± 0.4° F (± 0.2° C) at 20°C
Power
Voltage: 115/230 VAC 50/60Hz Current: 8Amp/4Amps

Model 284

SPECIFICATION

Display
½”x3½” digit LCD
Range : 200nC → Resolution : 0.1nC
Range : 20nC → Resolution : 0.01nC
Accuracy
2%
Output
0 to ±2 volt analog
Battery
9 volt Eveready #216 or equivalent NEDA #1604.
Battery life over 400 hours.
Dimensions
6”x3½”x21/8” (15 cm x 9 cm x 5.5 cm)
Weight
8½ oz. (0.24kg) with battery

제품 구성품

OPTIONAL

Faraday Cup Model 284/22A
Faraday Cup Model 284/22B
Model 1P20B Point Contact Probe

Model 4431T

SPECIFICATION

Power
Voltage: 90-260VAC, 50/60Hz; 0.75 Amps max
Input: IEC Socket with 6’ (2m) cable with NA plug (Std)
High Voltage Power Supply
Adjustable ±800-±1,200V
Discharge Network
STM11.31: 100pF & 1500Ω
Signal Output
Current output: Tektronix CT-1: 5mV/mA @ 500Ω:
Capacitive Sensor
Electrodes: 0.875” (22mm) stainless steel Dielectric: 0.5” (12mm) Capacitance: 6pF
Dimensions
4”W x 12”D x 6”H (102x305x152mm)

제품 구성품

PBT-531

SPECIFICATION

PDS-510 HBM Discharge Simulator
Discharge Output
Adjustable up to 4kV
Power Supply Output
Adjustable up to 4kV
Power Supply Accuracy
±2%
Discharge Waveform
(ANSI/ESD STM11.31)
Rise me @ 0Ω: < 10 nsec
@ 500Ω: < 20 nsec
Fall me @ 0Ω: 150±15 nsec
@ 500Ω: 200±20 nsec
Ringing: < 15%
Power
Batteries: 4 each AA 2600 mAh low self-discharge, Rechargeable, Nickel-Metal Hydride Battery
Power Supply: From 90 VAC to 264 VAC, 47 Hz to 63 Hz. UL and TUV 60950-1 approved
Battery Life
>100 hours (with average of 50 discharges in 8 hours)
Dimensions
(L x W x H) 7.5″ x 4.5″ x 2.5″ (19.05 x 11.43 x 6.35 cm)
Weight
2.0 lbs (907g)
PBF-520 Shielded Bag Fixture
Base Material
Flame Retardant Polycarbonate
Surface Resistivity
>1×1013Ω
Discharge and Ground Electrodes
1.5” (3.8 cm) brass, gold plated
Capacitive Probe
0.830” (2.1 cm) brass, gold plated
High Voltage Resistor
500 Ohms, 1% tolerance
Support Area Surrounding the Ground Electrode
8 x 10” (20 x 25 cm), without Base Extension attached
8 x 15” (20 x 35 cm), with Base Extension attached
Dimensions
(L x W x H) 11” x 8” x 4” (27.94 x 20.32 x 10.16 cm), without Base Extension
(L x W x H) 6” x 8” x 1” (15.24 x 20.32 x 2.54 cm), Base Extension
Weight
3.0 lbs (48 oz), without Base Extension
1.0 lbs (16 oz), Base Extension

제품 구성품

PGT-61-164

SPECIFICATION

Case Construction
ABS UL 94 V/0/5VA rated
Display Update For Volt
Less than 2.5 times second
Over-Range Indication on All Functions
Display “OL”
Dimensions
6.4” (L) x 3” (W) x 1.4” (D)
Weight
9.4oz (267g)

Measurement Specifications

MEASUREMENT
RANGES
RESOLUTION
ACCURACY
Line Voltage
85.0 – 250.0 VAC
0.1V
1.0% ± 2V
Peak Line Voltage
121.0 – 354.0 VAC
0.1V
1.0% ± 2V
Frequency
45.0 – 65.0 Hz
0.1Hz
1.0% ± 2V
% Voltage Drop
0.1% – 99.9%
0.1%
2.5% ± 2%
Voltage Loaded
10.0 – 250.0VAC
0.1V
2.5% ± 2%
Neutral-Ground V
0.0 – 10.0VAC
0.1V
2.5% ± 2%

제품 구성품

PGA-710B

SPECIFICATION

Size
5.0 in (12.7 cm) L x 2.8 in (7.1 cm) W x 1.35 in (3.4 cm) H
Weight
6.5 ounces (185 grams)
Input
< ± 2 Volts
OutPut
USB Cable
Sample Rate
50, 100 and 200 Samples per second
File Limit
During Sleep Mode (Main Power Battery Switch ON, unit OFF) : 8 – 10mA
Panel key pad is energized and unit is in standby mode.
During Normal Operations : 108mA
Unit is operating in remote or computer mode
Computer USB Battery Charge : 100mA
Operating in computer mode and receiving 100mA current from USB port reduces battery drain rate to 8mA
Battery Charge from AC/DC Charger : 280mA
Charge battery with Main ON, OFF, or during operations

제품 구성품

OPTIONAL

PFK-100 Electrostatic Field Meter Set

Measure Electrostatic Fields, Ionizer Balance and Body Voltage Testing

CVM-780

SPECIFICATION

Basic Description
High impedance contact voltmeter with active probe
DC Characteristics
Intended Operating Range: 0 to ±525 volts
Resolutions
1 volt, 3 digit display
Curcuit Accuracy
< ±1% (Note: Maximum measurement range is ±575 Volts for brief periods )
Dynamic Characteristics
Response Time: Approximately 1ms for a voltage step of 200V at the probe tip to the monitor output.
Digital display update: 3 readings per second.
Input Charge / Measurement
Approximately 100 femptocoulomb/volt (1.0×10-13 C/V)
Input Impedance
= 1.0xE14 ohm in parallel with < 1pF Capacitance
Recording Deviation
Typically < ±10 volts at ± 500 Volts
Probe Tips
Probe Tips Metal Nominally 0.95 and 0.45 inch Ceramic 0.90 and 0.40 inch ±0.05”
Power
4 each Energizer NH15-2500 Nickel-Metal Hydride rechargeable batteries (AA Size)
Measurement Tolerance
When measuring DC voltage reference
500 V Less or Equal to ±2 V
100 V Less or Equal to ±2 V
50 V Less or Equal to ±2 V
10 V Less or Equal to ±1 V
When measuring floating conductors
500 V Less or Equal to ±5% ±2 Counts
100 V Less or Equal to ±5% ±2 Counts
50 V Less or Equal to ±5% ±2 Counts
10 V Less or Equal to ±10% ±2 Counts

제품 구성품